Vitrek’s 944i
Hipot Tester combines AC dielectric strength testing (AC Hipot)
with DC dielectric strength testing (DC Hipot), Insulation
Resistance (HV IR) and a low Ohmic Earth Continuity measurement all in one feature-rich automated instrument.
This compact dielectric tester lets
you measure four electrical safety test parameters from a single
test connection-all with the touch of a button.
Our 944i
delivers top performance and versatility for manufacturing tests,
product safety engineering, component evaluation, as well as
research and development applications. Most hipot testers lack the
sensitivity to accurately detect low leakage current levels. The
944i uses dynamic ranging to provide leakage readings down to 2μA
full scale with 1nano-amp resolution.
When it comes
to overall dielectric performance - You'll find that the
multi-function 944i is loaded with features that even the larger
more expensive dedicated testers don't offer. Powerful features
including automatic calibration, built-in diagnostics, internal
"watchdog" safety monitor, user memory for up to 100 test routines,
arc detection and available interface package with IEEE-488.2, RS232
and direct test results printout capability.
Features/Benefits
-
AC & DC Hipot plus Insulation Resistance and
Ground Continuity in a single
feature rich, compact instrument
-
AC Hipot to 5KV standard, 7KV
and 10KVDC optional
-
Easy to use Full Menu Prompting, walks you
through the test set-up
-
Test Database stores 200 test steps for easy
recall
-
Powerful 35mA Sourcing capability
-
On-board Phase Detection measures "total",
"real" or "imaginary" Leakage Current
-
Continuous Output Frequencies from 50 to 100Hz
-
IR measurement to 65GΩ
-
Continuity measurement 8.5Ω full scale, resolution to 1mΩ
-
Leakage Current measurement to 1na
-
Optional IF-4 Interface package includes GPIB,
RS232, VICL & Printer Ports
-
Separate Ramp & Dwell, Min & Max current limits
-
Covers-on Automatic Calibration
-
Built-in controls for HV Scanner and Ground Bond
Tester
-
Mark Certified to EN61010
-
Includes NIST Traceable Certification
-
Three Year Extended Warranty
-
Made in the USA
Top of Page
The
intelligent choice for operator safety
At Vitrek,
operator safety is not just a concern, it is a relentless
commitment. So we engineered the 944i with a high-impact injection
molded front-end with no exposed metal parts to present a potential
user shock hazard. We put in an internal "watchdog" safety monitor
which automatically shuts down the output in the event of an
inadvertent disconnect. And the Vitrek 944i boasts an impressive
user programmable "Arc Detection" circuit, capable of shutting down
the high voltage system within microseconds.
For added peace
of mind the onboard diagnostics of the 944i execute a complete check
of all safety systems every time you power up.
If your
product requires one of these labels. . .

Demand that the
Vitrek 944i label is on your Dielectric Analyzer. Only Vitrek
minimizes overall test time and maximizes system reliability by
combining four dielectric test functions in one unit.
It would hardly
seem possible that a slow dielectric tester could back-up an entire
production line. Yet with today's imposing 100% dielectric
inspection requirements, the antiquated manual hipot tester is often
the manufacturing bottleneck. Imagine the consequences of the tester
breaking down or worse yet passing a defective product.
Just a few good
reasons why each 944i Multi-function Dielectric Analyzer is equipped
with two high performance microprocessors - to speed testing
throughput and to execute exhaustive onboard diagnostics. Simply
put, the Vitrek 944i provides the fastest, most accurate and
reliable test results available. And it does it for a surprisingly
modest price.
Top of Page
Powerful
yet easy to use
We've harnessed
the power of the microprocessor to make this dynamic instrument
simple and easy to operate. Specifically, the extensive use of menu
prompting walks you through the test setup. Built-in user memory
allows you to store, recall and edit up to 100 different test
programs. All test parameters are fully user selectable - no hidden
switches or pots to hamper operation.
The result is a
clean, streamlined panel design that demands noticeably less effort
to operate - and the unrivaled power to execute customized test
routines at the touch of a button.
Capacity
to excel
With its full
35mA sourcing capability, the Vitrek 944i possesses the horsepower
required to drive today's capacitively-filtered
circuits - loads which the 10mA to 25mA units just can't handle.
Yet, what sets the 944i apart from the rest is its ability to
discriminate between in-phase and out-of-phase leakage currents.
This advanced
feature enables you to measure the resistive current associated with
insulation breakdown, the reactive component or the combination of
both.
Vitrek Service & Support: Finest in the Instrument Industry
At Vitrek, we
manufacture automated test instruments that are human engineered for
friendly easy operation. But that's not where our commitment stops.
We back you up with a worldwide network of Vitrek support centers.
And we provide support by phone or email for any need or questions
that may arise.
This all adds up
to one point - you'll enjoy doing business with Vitrek. Our service
technicians, application engineers and customer service satisfaction
people just wouldn't have it any other way.
Top of Page
Specifications
AC DIELECTRIC MODE:
AC Voltage Output
Voltage uncertainty is equal to the square root of the selected
output. (V1/2)
|
Output |
Formula |
Accuracy |
|
100 VAC |
(100)1/2 |
± 10.0 VAC |
|
500 VAC |
(500)1/2 |
± 22.4 VAC |
|
1000 VAC |
(1000)1/2 |
± 31.6 VAC |
|
5000 VAC |
(5000)1/2 |
± 70.7 VAC |
Output Frequency:
50Hz to 100Hz with 1Hz resolution independent of line power
frequency.
Output Waveform:
Digitally synthesized, low distortion sine wave.
AC Current Sensing (AC Leakage)
|
Range |
Resolution |
Accuracy |
|
12.00uA |
10nA |
0.5% of reading ± 1uA |
|
1.200mA |
1uA |
0.5% of reading ± 10uA |
|
35.0mA |
100uA |
0.5% of reading ± 500uA |
Current Sensing:
Resistive current (in-phase),
reactive current (out-of-phase) or total current.
AC Current Limits
|
Range |
Current Limit |
|
50 to 5000 VAC |
35 mA |
DC DIELECTRIC
MODE:
DC Voltage Output
Voltage
uncertainty is equal to the square root of the selected output times
0.75. (V1/2 *0.75)
|
Output |
Formula |
Accuracy |
|
100 VDC |
(100)1/2
* 0.75 |
± 7.5 VDC |
|
500 VDC |
(500)1/2
* 0.75 |
± 16.8 VDC |
|
1000 VDC |
(1000)1/2
* 0.75 |
± 23.7 VDC |
|
5000 VDC |
(5000)1/2
* 0.75 |
± 53.0 VDC |
|
7000 VDC* |
(7000)1/2
* 0.75 |
± 62.7 VDC |
* Requires option DC-7
DC Current Sensing (DC Leakage)
|
Range |
Resolution |
Accuracy |
|
0 - 1720 ηA |
1ηA |
0.4% of reading ± 8 ηA |
|
1.720 μA - 17.30 μA |
0.01μA |
0.3% of reading ± 0.05μA |
|
17.30μA - 1.785mA |
0.001mA |
0.3% of reading ± 0.015 mA |
|
1.785 mA - 35.50 mA |
0.01 mA |
0.3% of reading ± 0.1 mA |
Voltage Coefficient:
15pA/V
DC Current Limits:
|
Range |
Current Limit |
|
50 to 5,000 VDC |
35.5mA |
|
5,000 to 7,000 VDC* |
19mA |
*
Requires DC-7 option
DC Voltage Output with Optional DC-10:
Voltage uncertainty is equal to the square
root of the selected output times 1.5.
(AC Voltage specs remain the same as
the standard 944 as described above)
Examples:
|
Output |
Formula |
Accuracy |
|
100 VDC |
100)1/2
* 1.5 |
± 15.0 VDC |
|
500 VDC |
(500)1/2
* 1.5 |
± 33.5 VDC |
|
1000 VDC |
(1000)1/2
* 1.5 |
± 47.4 VDC |
|
5000 VDC |
(5000)1/2
* 1.5 |
± 106.1 VDC |
|
7000 VDC* |
(7000)1/2
* 1.5 |
± 125.5 VDC |
|
10000 VDC* |
(10000)1/2
* 1.5 |
± 150.0 VDC |
DC Current Limits with option DC-10 :
(Output
power is limited to 165 Watts)
|
Range |
Current Limit |
|
50 to 10,000 VDC |
15.5mA |
|
10,001 to 14,000 VDC |
10.0 mA |
HIGH RESISTANCE MODE:
(Insulation Resistance)
(Firmware
limits to minimum of 1.9kOhms. Output power can not exceed 165
Watts for standard units, 135 Watts for CLT option and 130 Watts for
FR-4 mod).
|
Resistance Range
(guaranteed in spec) |
Minimum Resistance |
Test Voltage |
|
2.6GΩ |
1.9KΩ |
>= 50VDC |
|
13GΩ |
500KΩ |
>= 500VDC |
|
26GΩ |
1MΩ |
>= 1,000VDC |
|
65GΩ |
2.5MΩ |
>= 2,500VDC |
Accuracy:
Combined uncertainty of DC Voltage Output + Current Measurement
Accuracy
LOW RESISTANCE MODE:
(Earth Continuity)
|
Full
Scale |
10
Ω Maximum including lead resistance. |
|
Resolution |
1mΩ |
|
Test
Current |
50 mA |
|
Lead
Compensation |
Automatic offset circuit "nulls" out two wire lead
resistance |
|
Accuracy |
0.3% of
reading + 30mΩ (up to 5Ω) |
Top of Page
GENERAL
SPECIFICATIONS:
|
Arc
Detection Sensitivity |
User
selectable. Fast = 1 arc in a 100mS period. Medium = 15 arcs
in a 100mS period. Slow = 50 arcs in a 100mS period. |
|
Display |
40
character alphanumeric LCD |
|
High
Voltage Indicator |
Lightning bolt illuminated by independent output voltage
monitor. Audible alarm sounds prior to voltage output
unless disabled in firmware. |
|
Voltage Resolution |
Output
resolution is typically 3-4 volts AC or DC on standard
units. DC-10 and AC-10 outputs have a 6-8 volt resolution.
LVIR output resolution is 0.2 volts. Measurement resolution
is 1 Volt on all units, all options. |
|
Dwell Time Accuracy |
+2
seconds/minute, -500mS/minute |
|
Slew
Rate |
1
volt/second to 30000 volts/second. Firmware limits all ramps
to a minimum total time of 1 second. Actual ramp times vary
dependent on load. Ramp times are consistent when testing
similar devices. Test engineers should always observe the
test with an oscilloscope and adjust parameters to achieve
desired results with each specific device under test. |
|
Limits |
Test
automatically aborts when digitally stored minimum or
maximum limit is exceeded. Current and resistance minimums
will not trigger failures until >0.8 seconds of dwell time
has elapsed to allow for settling of DUT at dwell voltage. |
|
Remote Interface Option "IF-4" |
GPIB -
IEE-488.2 Subset compatible
RS-232C - 9600 Baud
Parallel - IBM/Centronics compatible
VICL - Vitrek Interface Control Loop (slaved to RS-232
interface) |
|
Accuracy |
Stated
specifications are bipolar and apply for one year within ±
5°C from calibration temperature (20 to 25°C). |
|
Temperature Coefficient |
0.1
times stated accuracy per °C beyond calibration temperature
window |
|
Calibration |
Traceable to the National Institute of Standards and
Technology (NIST). Automatic "Covers on" calibration,
internal or external. Calibration constants stored in
non-volatile memory. |
|
Warm
up Time |
No warm
up time is required for units stored in ambient temperatures
ranging from 64°F to 82°F (18°C to 28°C). A minimum 30
minute warm up time is recommended for units stored in
temperatures outside that range. A minimum 60 minute warm up
time is recommended prior to performing an external
calibration procedure. |
|
Operating Temperature |
64°F to
82°F (18°C to 28°C). Ensure units are well ventilated and
fan is operational at all times. Fan should be sucking fresh
cool clean air into unit for optimal performance.
Overheating and freezing may result in unexpected resetting
or lockup of units. Keep vents clear. |
|
Humidity |
Maximum
90% RH |
|
Power Consumption |
115/230
VAC ± 10%, 350 VA Max. Typical 40-50 Watts power
consumption in idle mode. Up to 300 Watts when sourcing
maximum output power to test device. Power consumption
varies with load during testing. |
|
Dimensions |
89mm H
x 432mm W x 457mm D (3.5" H x 17" W x 18" D) |
|
Weight |
14Kg
(34 lb.) Net / 18Kg (40 lb.) Shipping |
Top of Page
Prices listed in the table below are in US$ and apply to
North American destinations only. For
inter-continental product destinations, please
add 5% to the prices shown below.
| ID# |
Product |
Price $ |
Order |
| 944i |
Hipot Dielectric Tester |
$4,790.00 |
|
| IF-4 |
GPIB, Serial, VICL & Printer Ports |
795.00 |
|
| IF-1 |
Parallel Printer Interface Only |
495.00 |
|
| TL-10 |
Test Leads with High Voltage Probe |
95.00 |
|
| TL-20 |
Alligator Test Lead Set |
85.00 |
|
| TL-115A |
Test Leads with 115 VAC Outlet |
195.00 |
|
| TL-IEC |
IEC Power Socket Test Lead |
95.00 |
|
| RBJ-1 |
Remote Audible Alarm |
100.00 |
|
| RM-1 |
Rack Mount Kit |
75.00 |
|
| RSS-1 |
Remote Start Switch |
100.00 |
|
| RPO |
Rear Panel Outputs |
100.00 |
|
| SIJ-1 |
Safety Interlock Jack |
100.00 |
|
| PLC-2C |
PLC Logic Level Interface |
150.00 |
|
| QT-1 |
QuickTest™ 944i Software |
295.00 |
|
| QT-2 |
Bundled QuickTest™ Software with PC GPIB Card & Cable |
895.00 |
|
| QT-3 |
Quick Test™ Software for Win 98/2000/XP/NT includes NI-GPIB Card & Cable |
995.00 |
|
| LVIR |
Adds 10 to 99 VDC Operation for MW |
295.00 |
|
| LO-500 |
Extends Low W range to 500W |
295.00 |
|
| ADC-200K |
Automatic Load Discharge |
295.00 |
|
| GP-1 |
1 Meter Shielded GPIB Cable |
85.00 |
|
| GP-2 |
2 Meter Shielded GPIB Cable |
95.00 |
|
| RS-1 |
6' RS-232 Cable |
45.00 |
|
| PAR-1 |
6' Parallel Printer Cable |
45.00 |
|
| DC-7 |
Boosts Max DC Output to 7KV |
295.00 |
|
| DC-10 |
Boosts Max DC Output to 10KV |
1,295.00 |
|
| AC-10 |
Boosts Max AC Output to 10KV VAC |
1,255.00 |
|
| ACDC-10 |
10KVAC/10KVDC Range Option |
2,490.00 |
|
| OM-944 |
Additional Operator's Manual |
65.00 |
|
| SPK-944 |
5 Year Spare Parts Kit |
395.00 |
|
| EXW-2 |
2 Year Warranty, with initial purchase |
490.00 |
|
Top of PageI
Rental
Information
30 Day Minimum. Rates are for
standard models.
Additional options are available. Please contact factory for a quote
or for further assistance.
| Duration | Price | Billing Cycle | Buyout |
| 30 Days |
$725 | 30 Days | List Price minus 30% of Rental Paid |
| 60 Days | $1150/2 Months | 60 Days | List Price minus 40% of Rental Paid |
| 3 to 11 Months | $575/Month | 30 Days | List Price minus 50% of Rental Paid |
| 12 Months | $575/Month | 30 Days | $1 Buyout |
This document requires that you have the Adobe Acrobat reader which
you can download for free here.