950i Series Performance Specifications
DC Dielectric Tests
(DCW, DCIR, DCez, DCCAP)
DC Output Voltage: Â Â Â 2V to 6500V (951i & 952i)
                      2V to 11000V (953i,954i & 955i)
Resolution: Â Â Â Â Â Â Â Â Â Â 0.1V up to 999.9V, 1V above
Accuracy:             0.25% + 0.5V ( 1 year 23°C ± 3°)
Â
DC Current Sourcing: Â 50mA max, 25mA above 6000V (951i & 952i)
                      30mA max, 20mA above 6000V, 10mA above
                      7500V, 5mA above 9000V (953i, 954i, 955i)
Â
Ramp Time: Â Â Â Â Â Â Â Â Â Â 0.01 to 9999sec, 0.01sec resolution
                      or 0.1 to 50000V/sec, 0.1V/sec resolution
Â
Dwell Time: Â Â Â Â Â Â Â Â Â Â 0.02 to 9999 seconds or user terminated, 0.01sec
resolution
Â
Discharge: Â Â Â Â Â Â Â Â Â Â Â 25mA max in solid-state constant current discharge mode,
                      selectable same as ramp rate or ramp to next level if next
                      step is DCW,  DCez or DCIR
Â
DC Leakage Current: Â Â Measurement Range: 0 to +/-200mA,
Resolution: Â Â Â Â Â Â Â Â Â Â Â 4 digits (9999 counts) down to 100pico-amp
Accuracy:              0.25% + 0.5nA + ½ digit ( 1 year 23°C ± 3°)
Â
Min & Max Current limits: Â Â Â Â Selectable for Ramp and Dwell 100 pico-amps and up
Â
Measurement Period: Â Â 1 power line cycle (50/60Hz)
Â
Insulation Resistance (IR): Â Â Test modes include: End on pass reading, end
                           on fail or end on dwell timer
Â
|
Test
Voltage
|
5% Accuracy* Max resistance
|
10% Accuracy* Max resistance
|
20% Accuracy* Max resistance
|
|
500V
|
50GΩ
|
100GΩ
|
200GΩ
|
|
1000V
|
100GΩ
|
200GΩ
|
400GΩ
|
|
2500V
|
250GΩ
|
500GΩ
|
1TΩ
|
|
5000V
|
500GΩ
|
1TΩ
|
2TΩ
|
|
10000V
|
1TΩ
|
2TΩ
|
4TΩ
|
*Above uncertainties are approximate, IR accuracy is determined by adding
output voltage accuracy to current measurement accuracy in percentages.
Â
Â
AC Dielectric Tests
(ACW, ACIR, ACez, ACCAP)
Â
AC Output Voltage: Â Â Â 20V to 6000V RMS (951i, 952i, 953i, 954i)
                     50V to 10,000V RMS (955i)
                     100V to 30,000V RMS (option AC-30)
Â
Resolution: Â Â Â Â Â Â Â Â Â 0.1V up to 999.9V, 1V above
       Accuracy:            0.5% + 1.5V (+ 0.01% per Hz above 100Hz)
Decrease max output voltage by 0.1% per Hz, above 100Hz
Â
Decrease max voltage by 12.5V/mA loading (25V/ma 955i)
Â
Â
AC Current Sourcing:Â Â Â Â Â Â Â Â Â Â Â Â 50mA RMSÂ max (951i, 952i, 953i, 954i)
                          30mA RMS max , (955i)
                          200mA RMS with option AC-2 (2KVAC RMS max output)
                          10mA RMS Max with AC-30 option
Â
Â
Output Frequency:Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Digitally synthesized, low distortion sinewave
                          20Hz to 500Hz, standard or AC-2)
                          40Hz to 500Hz, (955i)
                           0.1% accuracy, 0.1Hz resolution (1Hz above 99.9Hz)
Â
Â
Ramp Time:Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â 0 to 9999sec, 0.01sec resolution
                          or 0.1 to 100000V/sec, 0.1V/sec resolution
Â
Dwell Time:Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â 0.02 to 9999 seconds or user terminated, 0.01sec resolution
Â
Accuracy: Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â 0.05% + 20mS
Â
Ramp Down:                          Selectable same as ramp time or ramp to next level if next test step is an ACV test type
AC Leakage Current: Â Â Â Â Â Â Measurement Range: 0 to +/-200mA RMS
        Resolution:                            4 digits (9999 counts) down to 100pico-amps
        Accuracy:                  0.5% + 10nA (add 0.005% per Hz above
                                 100Hz) Selectable min & max limits for Ramp & Dwell,
                                 from 100 pico amps up Measurement Period: 1 power
                                 line cycle (50/60Hz)
Phase Measurement:Â Â Â Â Â Â Â Â Â Â Â Total RMS current, In-phase current, Quadrature
                           current (reactive/out-of-phase)
Accuracy:                  0.01° per Hz, relative to output voltage
AC Insulation Resistance:Â Â Â Â Test modes include: End on pass reading, end on
                           fail or end on dwell timer
|
Test Voltage
(50/60Hz)
|
10% Accuracy* Max resistance
|
20% Accuracy* Max resistance
|
|
500V
|
5GΩ
|
10GΩ
|
|
1000V
|
10GΩ
|
20GΩ
|
|
2500V
|
25GΩ
|
50GΩ
|
|
5000V
|
50GΩ
|
100GΩ
|
|
10000V (955i only)
|
100GΩ
|
200GΩ
|
*Approximate uncertainties, ACIR accuracy is determined by adding
output voltage accuracy to in-phase current accuracy in percentages.
Â
Â
Ground Bond Tests
(GB, GBez - 952i, 954i, 959i)
Â
Test Current: Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â 0.1 to 40A RMS, 0.001A resolution
Â
Accuracy: Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â 0.5% + 5mA accuracy
                          (add 0.005% per Hz above 100Hz)
Â
Test Frequency: Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â 40Hz to 500Hz
Â
Resolution: Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â 0.1Hz (1Hz above 99.9Hz) Accuracy: 0.1% accuracy
Â
        Waveform:                Digitally synthesized, low distortion sinewave
Measurement Configuration:Â 4-Terminal Kelvin
        Compliance Voltage:             6.5V RMS, may be user limited to a lower level with
                                  0.01V resolution
Resistance Range:                6.5 ohms at 1A decreasing to 162.5 milli-ohms max at 40A
Â
Max load impedance: Â Â Â Â Â Â 10 ohms
Â
Ramp Time: Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â 0 to 9999sec, 0.01sec resolution
Accuracy: Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â (1% + 0.1sec) accuracy
Dwell Time: Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â 0.02 to 9999sec or user terminated, 0.01sec resolution
Accuracy: Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â 0.05% + 20ms
Ramp Down: Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Fast (35mS), same rate as ramp up or may be
                           skipped if next test step is a ground bond test
Voltage Sense:Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Range: 0 to 8 v rms
       Resolution:                             4 digits down to 10uV Accuracy: 0.5% + 30uV
Phase Measurement:Â Â Â Â Â Â Â Â Â Â Â Â RMS, In-phase and Quadrature measurements
                           0.01° per Hz phase relative to test current
Â
Low Resistance Measurement
Â
       Measurement Range:                 0 to 150Kohm (999.9mΩ to 99.99KΩ, 149.9KΩ in 7 ranges)
Resolution:                              4 digits, 100µΩ on 1ohm range
Â
Test Current:                            55mAdc constant current up to ≈ 91Ω,
5VDC constant V above
Â
Accuracy (4-wire):                    0.5% + 0.002ohm + ½ digit up to 30K ohm
                           1.5% + ½ digit above 30K ohm
                           5% + 1 digit from 100K to 150K ohm
                           Add 20mΩ for 2 -wire
Â
Line Leakage Current & Voltage Measurement
(Models 951i - 955i only)
Â
Voltage Measurement:            0 to ± 8KVDC 6KV RMS AC (951i & 952i)
                           0 to ± 11KVDC 8KV RMS AC (953i & 954i)
                            0 to ± 11KVDC 10KV RMS AC ( 955i)
Â
Resolution: Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â 0.1V, 1 V above 999.9V
DC Accuracy: Â Â Â Â Â Â Â Â Â Â Â Â Â Â 0.25% + 0.5V
AC Accuracy: Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â 0.5% + 1.5V
Â
Leakage Current:                   0 to ± 200mA DC or RMS AC
Resolution: Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â 4 digits (9999 counts) down to100pico-amps
Â
DC Accuracy: Â Â Â Â Â Â Â Â Â Â Â Â Â 0.25% + 0.5nA
AC Accuracy: Â Â Â Â Â Â Â Â Â Â Â Â Â Â 0.5% + 20nA
Test Results                          Test Time: 0.02 to 9999 sec Last, Minimum, Average
                          & Max V& A reading plus arc current
Â
Pulse Mode (Flash) Test
(Option PMT-1 available on models 951i, 952i, 953i, 954i)
Â
Test Waveform: Â Â Â Â Â Â Â Â Â Â Â Trapezoidal (Selectable positive polarity, negative
                          polarity or bi-polar)
Â
Ramp Up/Down Time:Â Â Â Â Â Â Â Â Â Â Â 1ms (0.5ms for option AC2) to 30mS with 0.1ms
                          resolution
Â
Dwell Time:Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â 1ms (0.5ms for option AC2) to 30mS with 0.1ms
resolution
Â
Test Voltage:Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â 50V to 8000V (20V to 2750V with option AC-2)
Resolution:Â 0.1V up to 999.9V, 1V above
Accuracy: 0.25% + 1.5V
Â
General Specifications
Test Lead Safety Sense:Â Â Â Â TLSSâ„¢ Technology continuously verifies that test leads are
                          properly connected prior to and during HV, 4-wire Low Ohms
                          and Ground Bond testing
Â
Arc Detection Sensitivity:Â Â Â Â User selectable 3 microseconds to 30 microseconds.
Dwell Time Accuracy: Â Â Â Â Â Accuracy: 0.05% + 20mS
RS232 Interface:           Digital I/O,     VICL: Vitrek auxiliary unit control loop
                          Remote Interface Option "IF-3":
                          GPIB - IEEE-488.2 Subset compatible
                          RS232C - 1200, 2400, 4800, 9600 Baud rate selectable
                          Parallel - IBM/Centronics compatible
Â
Accuracy: Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Stated specifications are bipolar and apply for one year
                        within 5°C from calibration temperature (20 to 25°C).
Â
Operating Temperature:       0 C to 50° C
Â
Humidity:                           90% RH, 0 to 40° C
Â
Power:Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â 115/230 VAC + 10%, 350 VA Max
Â
Dimensions: Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â 89mm H x 432mm W x 457mm D (3.5" H x 17" W x 18" D)
Â
Weight:Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â Â 9Kg (18 lb.) Net / 18Kg (25 lb.) shipping (951i, 953i, 959i)
                       14Kg (28 lb.) Net / 18Kg (35 lb.) shipping (952, 954i, 955i)