948/964i Hipot High Voltage Switching System with GPIB

The Solution to Multi-point Automated Hipot Dielectric Testing

Have a question about the 948i Switching System? Ask An Expert!

 

Unlimited Scanning Possibilities with 7KV, 10KV, 15KV, 40A, & 70A Switching Modules

When a single 948/964i Hipot High Voltage Switching System is combined with a Vitrek automated Hipot tester, up to 48 test points can be scanned for high voltage isolation - multiple units work together to increase your test point count. Combine two 948i's for 96 points of switching capability or ten units for 480 point capacity. Expansion possibilities are virtually unlimited. And with voltage switching capability up to 15,000 volts or current switching as high as 70 amps - the 948i can handle just about any test requirement that comes along.

Modular Configuration Assures that Your Requirements are Satisfied

The 948i is easily configured to meet your specific test needs. For starters, select the voltage rating and number of HV switching channels desired.  In addition to hipot and IR - do you also need to do continuity or low resistance? How about LCR measurement or 4-wire resistance?  The MX modules can be added or customized to allow additional measurement equipment to access the matrix.

In a typical application each test point needs to be bi-polar (high for some tests and low for others), multiply the number of test points in the DUT by two and that is how many relay channels you will need.

A single 948i can accommodate 24 bi-polar test points with three 8 channel cards connected to the high voltage bus and with three 8 channel cards connected to the return bus. To add continuity or low resistance measurement capability to the system, an MX-2 card will allow you to use the matrix for either hipot or low  resistance testing.  For higher test point counts additional 948i's can be connected to accomplish the task.  If you have any questions about configuring one or more 948i's Vitrek applications assistance is just a phone call or a click away.

 

 

Need help configuring your 948i? Ask An Expert!

Set-Up is a Snap

 

The 948i is fast and easy to use so there is no excuse for using the hit-and-miss manual probing method of high voltage multi-point Hipot testing. The 948i increases operator safety by reducing exposure to high test voltages while enhancing product quality and reliability. Consistent and repeatable hipot test results are automatically documented making it simple to prove compliance with ISO 9000 quality standards.

Setting up the switching pattern for the 948i is a snap. The standard GPIB interface port allows the system controller to download, verify and activate any desired switching profile you can dream up. Or if you prefer, an optional RS232 interface is also available. And if programming code is not your ideal way of spending the day, Vitrek QuickTest graphical software can get you up and running in no time at all. For those who are looking for a stand alone system the 948i can be controlled directly by the Vitrek 944i in a PC free environment. So, no matter how you like it - Vitrek puts you in control of your automated high voltage test requirement.

Now that there are no more excuses, isn't it about time you automated your high voltage test application?


Voltage Switching Ratings to 7KV, 10KV or 15KV

          Current switching to 40 or 70 Amps

          Reduces operator exposure to high voltage

          Improves test quality, consistency and repeatability

          Enhances end product quality and reliability

          Scans up to 48 points with single compact unit

          GPIB Standard - RS232 no extra charge

          May be controlled directly by Vitrek Hipot Tester

          Available QuickTest™ graphical programming software

          CE mark certified to EN 61010

          Made in the USA

 

The below is a schematic representation of two 8 channel scanning modules. One is connected to the high voltage output of the tester and the other to the return sense. This configuration provides bi-polar test capability for the 8 test points. Additional cards and units can be added for greater capacity.

948i Switch Card Diagram

948i High Voltage Scanning System (7 Card Capacity) $2,490.00

HV7-8B 7KV Single Input Bus 8 Output Relays Terminated in Black HV Term $800.00

HV7-8R 7KV Single Input Bus 8 Output Relays Terminated in Red HV Termin $800.00

HV7-4R 7KV Single Input Bus 4 Output Relays Terminated in Red HV Termin $500.00

HV7-4B 7KV Single Input Bus 4 Output Relays Terminated in Black HV Term $500.00

HV7-MX2 Dual Input Single Output MUX Card (Selectable HV or Continuity C $400.00

HV10-8B 10KV Single Input Bus 8 Output Relays Terminated in Black HV Ter $1,000.00

HV10-8R 10KV Single Input Bus 8 Output Relays Terminated in Red HV Termi $1,000.00

HV10-4B 10KV Single Input Bus 4 Output Relays Terminated in Black HV Ter $600.00

HV10-4R 10KV Single Input Bus 4 Output Relays Terminated in Red HV Termi $600.00

HV10-MX2 Dual Input Single Output MUX Card (Selectable HV or Continuity C $500.00

HV15-8B 15KV Single Input Bus 8 Output Relays Terminated in Black HV Ter $1,200.00

HV15-8R 15KV Single Input Bus 8 Output Relays Terminated in Red HV Termi $1,200.00

HV15-4B 15KV Single Input Bus 4 Output Relays Terminated in Black HV Ter $700.00

HV15-4R 15KV Single Input Bus 4 Output Relays Terminated in Red HV Termi $700.00

HV15-MX2 Dual Input Single Output MUX Card (Selectable HV or Continuity C $600.00

948i-QT-1 QuickTest Software $295.00

GP-1-948 1 Meter Shielded GPIB Cable $85.00

948i-GP-2 2 Meter Shielded GPIB Cable $95.00

948i-RS-1 6' RS-232 Cable $45.00

HC-7 4 Channel 70A Source & 4 Ch Sense Relay Card $700.00

HC-4 4 Channel 30A Source & Sense Relay Card $500.00

MC-1L HV Mating Connector with 4' HV Lead $25.00

948i-RM-1 Rack Mount Kit $75.00

964i HV/HC Switching System 8 card capacity $2,490.00

OM-948 Additional Operator's Manual $35.00

SPK-948 5 Year Spare Parts Kit $395.00

Switch Card Specifications

HV7 Series - 7KV High Voltage Switch Card

  
Maximum Switching Voltage 7kVDC, 5kVAC RMS 50-60Hz
Maximum Switching Current 2A DC or RMS AC
Maximum Switching Power 50 Watts
Relay Switching Time 4mS (from applied coil current)
Relay Configuration Form A, SPST Normally Open

HV7 Series Switch Card Configuration:

HV7-8R 7KV Single input bus 8 output relays terminated in Red HV terminals,

HV7-8B 7KV Single input bus 8 output relays terminated in Black HV terminals

HV7-4R 7KV Single input bus 4 output relays terminated in Red HV terminals

HV7-4B 7KV Single input bus 4 output relays terminated in Black HV terminals

HV7-MX2 Dual input Single output MUX Card (selectable HV or continuity capability)

 

HV10 Series - 10kV High Voltage Switch Card

Maximum Switching Voltage 11kVDC, 8kVAC RMS 50-60Hz
Maximum Switching Current 2A DC or RMS AC
Maximum Switching Power 50 Watts
Relay Switching Time 4mS including debounce
Relay Configuration Form A, SPST Normally open

HV10 Series Switch Card Configuration:

HV10-8R 10KV Single input bus 8 output relays terminated in Red HV terminals,

HV10-8B 10KV Single input bus 8 output relays terminated in Black HV terminals

HV10-4R 10KV Single input bus 4 output relays terminated in Red HV terminals

HV10-4B 10KV Single input bus 4 output relays terminated in Black HV terminals

HV10-MX2 Dual input Single output MUX Card (selectable HV or continuity capability)

 

HV15 Series High Voltage Switch Card:

Maximum Switching Voltage 15kVDC, 10kVAC RMS 50-60Hz
Maximum Switching Current 2A DC or RMS AC
Maximum Switching Power 50 Watts
Relay Switching Time 4mS including debounce
Relay Configuration Form A, SPST Normally open

HV15 Series Switch Card Configuration:

HV15-8R 15KV Single input bus 8 output relays terminated in Red HV terminals,

HV15-8B 15KV Single input bus 8 output relays terminated in Black HV terminals

HV15-4R 15KV Single input bus 4 output relays terminated in Red HV terminals

HV15-4B 15KV Single input bus 4 output relays terminated in Black HV terminals

HV15-MX2 Dual input Single output MUX Card (selectable HV or continuity capability)

HC-4 40 Amp High Current Source & Sense Switch Card

 

Maximum Input Voltage 700 VDC, 50 VAC RMS
Maximum Switching Voltage 277VAC, 30VDC
Maximum Current 40A DC or RMS AC
Relay Switching Power 500 Watts
Card Configuration 4 channels of high current relays and 4 channels of sense relays
Relay Configuration All Form A, SPST Normally open

HC-4 High Current Switch Card Configuration:

HC-4R 4 HC Source & 4 sense relays terminated in Red Terminals.

HC-4B 4 HC Source & 4 sense relays terminated in Black Terminals.

 

HC-7 70 Amp High Current Source & Sense Switch Card

Maximum Input Voltage 700 VDC, 500 VAC RMS
Maximum Switching Voltage 277VAC, 30VDC
Maximum Current 70A DC or RMS AC
Relay Switching Power 500 Watts
Card Configuration 4 channels of high current relays and 4 channels of sense relays
Relay Configuration All Form A, SPST Normally open

HC-7 High Current Switch Card Configuration:

HC-7R 4 HC Source & 4 sense relays terminated in Red Terminals.

HC-7B 4 HC Source & 4 sense relays terminated in Black Terminals.

 

General Specifications

Switch Card Capacity Holds up to 7 switch cards
High Voltage Terminals All HV terminals are 15KV rated and supplied with mating connector, optional HV leads sets available on request
Remote Interface GPIB standard, RS232 optional
Indicators L.E.D. status indicators for Test, Standby, Active, Talk, Listen, Address
Diagnostics Built-in self diagnostics test ROM and RAM 
Operating Temperature 0°C to 50°C
Humidity 80% RH 0°C to 40°C
Power 115/230 VAC +/- 10% VA max
Dimensions 89mmH x 432mmW x 485mmD (3.5"H x 17"W x 19"D)
Weight 12Kg (27 lb.) Net/16Kg (36 lb.) Shipping
Standard Accessories Operator's manual, hipot connection test leads and power cord
Warranty 1 year parts and labor
Safety CE mark certified to EN61010
30 Day Minimum.  Rates are for standard models. 
Additional options are available. Please contact factory for a quote or for further assistance.
Duration Price Billing Cycle Buyout
30 Days $400 30 Days List Price minus 30% of Rental Paid
60 Days $600/2 Months 60 Days List Price minus 40% of Rental Paid
3 to 11 Months $300/Month 30 Days List Price minus 50% of Rental Paid
12 Months $300/Month 30 Days $1 Buyout

In order to better assist you in the future, we would like to get some basic information from you before you download any of our Manuals. We respect your right to privacy and do not and will not ever share this information with any third party. Please review our privacy policy for more information.

Country:*
Name:*
Company:
Phone (optional):
E-mail:*
Please subscribe me to the Vitrek newsletter mailing list.