Everything You
Need In a Multi-point Safety Analyzer for an Exceptionally Low Price
The V5 Sets the Standard for Versatility and Value. Only Vitrek, the technology leader in electrical safety test automation, could build a feature rich, full function 5-in-1 Hipot Electrical Safety Tester with such high performance this high and a price this low. The new Vitrek V5 combines AC Hipot to 5KV, DC Hipot to 6KV, Insulation Resistance to 10 gig-ohms and high current Ground Bond measurement up to 42 Amps and a built-in 8 channel HV Scanner. And all this comes in one systems ready, low profile, rugged, rack mount package – complete with rear terminal outputs. Features/Benefits Built-in 8 Channel Scanner Tests AC, DC Hipot, IR & Ground Bond to UL, CSA, IEC & TUV Specs Up to 5KVAC 40mA Up to 6KVDC 10mA Multi-mode Insulation Resistance IR 50-1000V up to 10,000MΩ Ground Bond continuity up to 42Amps Leakage current resolution down to 10μA Fully programmable RS232 optional GPIB, IEEE-488.2, interface Selectable arc detection Available Safety Fault Interrupt (SFI)™ circuit CE mark certified EN61010 10 position test sequence memory 3 year extended warranty*
Top of Page Specifications AC Hipot
| Voltage
Range: | 100 to 5000 VAC RMS |
| Voltage
Setting Resolution: | 5V/step |
| Voltage
Regulation: | 1% of reading + 5V |
| Voltage
Accuracy: | 1% of reading + 10V |
| Voltage
Monitoring: | At Output Terminals |
| Output
Current: | 30mA to 40mA 180S max, 0 to 30mA
continuous. Below 500V 10mA max |
| Min/Max
Current Limits: | 0.1 to 40mA |
| Current
Accuracy: | 1% of reading + 50mA | Top of Page DC Hipot
| Voltage
Range: | 100 to 6000 VDC |
| Voltage
Setting | Resolution: 5V/step |
| Voltage
Regulation: | 1% of reading + 5V |
| Voltage
Accuracy: | 1% of reading + 10V |
| Voltage
Monitoring: | At Output Terminals |
| Max Output
Current: |
10mA continuous, below 500VDC 2mA
continuous |
| Min/Max
Current Limits: | 0.01 to 10mA |
| Current
Accuracy: | 1% of reading + 50mA | Insulation Resistance
| DC Output
Voltage: | 50, 100, 500 or 1000V |
| Resistance
Range: (Full Scale) | 50MW | 200MW | 9999MW |
| Resistance Range: | 10KW | 100KW | 1MW |
|
Resistance
Accuracy: (% of reading) |
| 50, 100 VDC | 1-50MW 5%
51-200MW 10%
201-1999MW 20% |
| 500, 1000 VDC | 1-500MW 5%
501-2000MW 10%
2001-9999MW 20% |
| Multi-Mode IR™: | Select 1 of 3 modes - test passes
on good reading, test fails on bad reading, pass/fail determined
at end of test time. | Top of Page Ground Bond/Continuity
| Test
Current: | 3A to 42 Amps RMS |
| Terminal
Configuration: | 4 wire Kelvin |
| Resistance
Range: | 0 to 600mW |
| Resistance
Resolution: | 100mW |
| Resistance
Accuracy: | 1% reading ± 3mW |
| Compliance
Voltage: | 8 volts at terminals |
| Current
Resolution: | 0.01A |
| Voltage
Resolution: | 0.01V | Top of Page General
| Test
Database: | 10 test sequences with up to 16
steps per test sequence |
| Front Panel
Lockout: | Prevents database edit while in
lockout mode |
| Arc Detect: | 10 level trip select |
| Remote
Interfaces: | RS232 -
Fully interactive PLC - Inputs; Start, Reset Outputs: Testing, Pass, Fail GPIB - Option GP-V5 |
| Timer
Accuracy: | 0.25% ± 50mS |
| Power
Source: | AC100v, 120V, 220V, 230V (±10%
50/60Hz), 350W max |
| Operating
Environment: | Indoor use, altitude up to 6500
ft (2000m). Installation Category II. Pollution Degree 2. |
| Operating
Temperature: | 0°C to 40°C, 80% RH |
| Storage
Temperature: | -10°C to 70°C, 70% RH |
| Dimension: | 17 W x 17 D x 5.25 H,
(432mm x 432mm x 133mm) |
| Weight: | 39lbs (18kg) net, 45 lbs (20kg)
ship |
| Accessories: | NIST Cal Cert, TL-20-60 &
K-2 Test
Leads, Instruction Manual and Power Cord. |
| Accuracy: | Stated specifications apply for 1
year at 23°C ± 5°C. |
| Calibration: | ANSI/NCSL Z540 Certificate of
Calibration with Data traceable to NIST. |
| Compliance: | CE Mark Certified to EN61010 and
EMC Directives EN50081-1 & EN50082-2. |
| Warranty: | * 3 Year Extended Warranty with
registration and annual factory calibration, 1 year parts and labor
standard. | QuickTest™ Software
| Option
QT-V5: | Bundled Test Automation Software
and PC Serial Interface Cable. Easy to load graphical test
sequence development and operating environment. Provides
time/date, operator, test parameters, test results and readings.
Results may be viewed, printed out or exported to a database
program. | Top of Page Prices listed in the table are in US$ and apply to North American destinations only. For inter-continental destinations, please add 5% to the prices shown below.
| ID# | Product | Price $ | Order |
| V5 | Hipot Electrical Safety Tester, Ground Bond Tester | $4,990.00 | | |
| GP-V5 | GPIB Interface Option |
495.00 | |
| TL-115-5 | Hipot/GB Outlet Test Adaptor | 195.00 | |
| TL-10-60 | High Voltage Probe Lead Set | 95.00 | |
| TL-20-60 | High Voltage Alligator Lead Set | 85.00 | |
| TL-IEC5 | IEC Socket Lead Set | 115.00 | |
| K-2 | High Current, 4-wire, 2 Clip Kelvin
Lead Set | 185.00 | |
| K-4 | High Current, 4-wire, 4 Clip Kelvin
Lead Set | 195.00 | |
| SFI-2 | Safety Fault Interrupt Option | 100.00 | |
| MHL-4U | Mating HV Lead 4 ft. unterminated | 45.00 | |
| MHL-4A | Mating HV Lead with Alligator Clip | 50.00 | | Top of Page Rental
Information 30 Day Minimum. Rates are for
standard models.
Additional options are available. Please contact factory for a quote
or for further assistance.
| Duration | Price | Billing Cycle | Buyout | | 30 Days |
$800 | 30 Days | List Price minus 30% of Rental Paid | | 60 Days | $1200/2 Months | 60 Days | List Price minus 40% of Rental Paid | | 3 to 11 Months | $600/Month | 30 Days | List Price minus 50% of Rental Paid | | 12 Months | $600/Month | 30 Days | $1 Buyout |

This document requires that you have the Adobe Acrobat reader which
you can download for free here. |